Long-term astigmatism after intraocular pressure sensor implantation and nonpenetrating glaucoma surgery: EYEMATE-SC
Colya N Englisch1, Karl T Boden, Peter Szurman
1From the Sulzbach Eye Clinic, Sulzbach/Saar, Germany (Englisch, Boden, Szurman, Schütte, Messias, Wakili); Department of Experimental Ophthalmology, Saarland University Medical Center, Homburg/Saar, Germany (Englisch, Langenbucher); Swiss Visio Glaucoma Research Centre, Montchoisi Clinic, Lausanne, Switzerland (Mansouri); Department of Ophthalmology, University of Colorado, Denver, Colorado (Mansouri); Knappschaftskrankenhaus Eye Clinic, University Hospital of the Ruhr University Bochum, Bochum, Germany (Dick); Ophthalmology, University Medical Centre of the Johannes Gutenberg University Mainz, Mainz, Germany (Hoffmann); Department of Ophthalmology, LMU University Hospital, LMU Munich, Munich, Germany (Mackert).
Purpose:
To investigate long-term astigmatism after combined nonpenetrating glaucoma surgery (NPGS) and implantation of the first miniaturized suprachoroidal intraocular pressure (IOP) sensor EYEMATE-SC.
Setting:
The study was conducted in 5 medical centers in 2 different countries.
Design:
Retrospective multicenter clinical study.
Methods:
Astigmatism of patients instrumented with the EYEMATE-SC IOP sensor was assessed over a follow-up period of 3 years. Refraction and corrected distance visual acuity (CDVA) were obtained preoperatively, after 6 months, 1, 2, and 3 years. A canaloplasty-operated patient cohort served as control. Astigmatism was evaluated using 3-dimensional power vector analysis involving the spherical equivalent M, and the Jackson crossed cylinder projections J 0 and J 45 . Exclusion criteria included neovascular and angle-closure glaucoma, myopia, axial length outside 22 to 26 mm, other ocular diseases, prior glaucoma surgery, other ocular surgery within 6 months (cataract surgery within 3 months) before NPGS, serious generalized conditions, and other active medical head/neck implants.
Results:
Multivariate analysis indicated no changes in astigmatism along the observation period in both the EYEMATE-SC (n = 24) and the canaloplasty (n = 24) group ( P > .05 or nonsignificant after Bonferroni correction). Astigmatism was unchanged between the EYEMATE-SC and the canaloplasty group at all timepoints ( P > .05). CDVA did not change along the observation period of 3 years in each of both groups ( P > .05).
Conclusions:
Despite its suprachoroidal localization, this study indicated that the miniaturized EYEMATE-SC IOP sensor did not negatively affect long-term astigmatism after combined implantation with NPGS.