Left bundle branch area pacing using a lumenless lead: A systematic literature review and meta-analysis
Kenneth A Ellenbogen1, Dedra H Fagan2, Patrick Zimmerman2
1Department of Cardiac Electrophysiology, Virginia Commonwealth University Health System, Richmond, Virginia, USA.
Background:
Although left bundle branch area pacing (LBBAP) has been shown to be a feasible option for delivering physiological pacing, data are largely limited to single-center reports. The aim of this analysis was to systematically assess the safety and efficacy of LBBAP with the Model 3830 lead among primarily bradycardia patients.
Methods And Results:
PubMed, Embase, Cochrane Library, and Google Scholar were searched for full-text articles on LBBAP using the SelectSecure Model 3830 lumenless lead. Rates and means were estimated using random- and mixed-effects models. Of 3395 articles, 53 met inclusion criteria, representing 6061 patients undergoing an implant attempt. Average patient age was 68.1 years (95% CI: 66.6, 69.6) and 53.1% were male (95% CI: 50.5%, 55.7%). The average implant success rate among bradycardia-indicated patients was 92.7% (95% CI: 89.5%, 94.9%). The overall estimated procedural adverse event rate was 2.5% (95% CI: 1.1%, 5.4%). The estimated septal perforation rate at implant was 1.6% (95% CI: 1.0%, 2.6%) with no adverse clinical sequelae reported. Pacing thresholds were low at implant (0.67 V [95% CI: 0.64, 0.70]) and remained stable through 12 months (0.76 V [95% CI: 0.72, 0.80]). Among bradycardia-indicated patients, LVEF remained stable from baseline to post-implant (59.5% [95% CI: 57.9%, 61.1%] vs. 60.1% [95% CI: 58.5%, 61.7%]).
Conclusion:
This meta-analysis including 6061 patients implanted with a Model 3830 lead for LBBAP found an average implant success rate of 92.7% and a procedural adverse event rate of 2.5% with stable electrical parameters and LVEF post-implant.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Extraction: Partition and Distribution Coefficients
For extracting a solute from an aqueous phase into an organic...
¹H NMR: Complex Splitting
Splitting diagrams or splitting tree diagrams are routinely used to depict such complex couplings. While drawing splitting diagrams, the splitting with the larger coupling constant is usually applied first.
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations
¹H NMR Signal Multiplicity: Splitting Patterns
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
IR Frequency Region: Fingerprint Region
The...