Commutable Blood Materials from the Fixed-Cell Method for Performance Evaluation of Blood Glucose by a Glucose Meter
Napaporn Apiratmateekul1,2, Jintana Nammoonnoy3, Gerald J Kost4
1Department of Medical Technology, Faculty of Allied Health Sciences, Phitsanulok 65000, Thailand.
Abstract:
Glucose meters provide a rapid blood glucose status for evidence-based diagnosis, monitoring, and treatment of diabetes mellitus. We aimed to evaluate the commutability of processed blood materials (PBMs) and their use in the performance evaluation of glucose meters. Two PBMs obtained by the fixed-cell method were analyzed for homogeneity, stability, and commutability. The compatibility of ten pairs between mass spectrometry and each glucose meter was categorized as compatible (mean paired difference ≤ 5%) and incompatible (mean paired difference > 5%). The performance of glucose meter 1 (n = 767) and glucose meter 2 (n = 266) was assessed. The glucose in the PBMs remained homogenized and stable for at least 180 days. Six out of ten pairs had commutable PBMs. Commutability of PBMs was observed in both well-compatible and incompatible glucose results. Target glucose values from mass spectrometry were significantly different (p ≤ 0.05) from consensus values in one group of glucose meters. When commutable PBMs were used, glucose meter 1 showed better performance than glucose meter 2, and the percentage of satisfaction was associated when using target values for glucose from mass spectrometry and consensus values, but the performance of glucose meter 2 was not associated. PBM from a fixed-cell method could be mass produced with acceptable homogeneity and stability. Commutability testing of PBMs is required prior to use in the performance evaluation of glucose meters, as the commutability of glucose in the PBMs obtained by a fixed-cell method was variable and depended on the individual glucose meter.
More Related Videos
07:55A Multi-Parametric Islet Perifusion System within a Microfluidic Perifusion Device
Published on: January 26, 2010
07:35Author Spotlight: Investigating the Blood Glucose Homeostasis in Murine Brain Using a Cost-Effective Hyperglycemic And Hypoglycemic Clamp Technique
Published on: January 26, 2024