Wideband Current Transducer Traceable Calibration up to 10 A and 1 MHz
Mohamed Ouameur1, Daniela Istrate1, François Ziade1
1Laboratoire National de Métrologie et d'Essais, 78197 Trappes, France.
Sensors (Basel, Switzerland)
|April 27, 2024
Summary
This study introduces an improved calibration method for current transducers, enabling accurate measurements up to 1 MHz and 10 A. This enhances energy efficiency management by providing traceable, high-frequency current data.
Area of Science:
- Electrical Engineering
- Metrology
Background:
- Energy efficiency is critical in industry due to rising electrical energy consumption.
- Accurate power quality and traceable metering are essential for energy efficiency management.
- Current transducers measure electrical events up to megahertz frequencies, including pulse and transient currents.
Purpose of the Study:
- To present an enhanced traceable calibration methodology for current transducers.
- To extend calibration capabilities to 10 A and 1 MHz, a tenfold increase in current level at high frequencies.
- To improve the accuracy and traceability of current transducer measurements.
Main Methods:
- Development of a traceable calibration methodology based on a metrological traceability chain.
- Utilizing a calculable current shunt for calibration.
- Extending calibration capabilities to 10 A and 1 MHz.
Main Results:
- Achieved traceable calibration of current transducers up to 10 A and 1 MHz.
- Established a calibration method that is unique for these parameters.
- Obtained an uncertainty of less than 0.2% for the transimpedance ratio.
Conclusions:
- The enhanced methodology significantly advances the calibration of current transducers at high frequencies and current levels.
- This improved calibration supports more effective energy efficiency management and power quality monitoring.
- The method provides a traceable and highly accurate solution for current transducer characterization.
Related Concept Videos
Instrument Calibration
178
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
178
Controlled-Current Coulometry: Overview
199
Controlled current coulometry, also known as amperostatic coulometry, is a technique used in electrochemical analysis to measure the quantity of a substance through the controlled passage of current. It involves the application of a constant current to an electrochemical cell containing the analyte of interest. As the current flows through the cell, the analyte undergoes a redox reaction at the electrode surface, resulting in a charge transfer. By monitoring the time required for a certain...
199
Galvanometer
2.1K
Common devices, including car instrument panels, battery chargers, and inexpensive electrical instruments, measure potential difference (voltage), current, or resistance using a d'Arsonval galvanometer. This electromechanical instrument is also known as a moving coil galvanometer.
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform...
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform...
2.1K
Design Example: Strain Gauge Bridge or Wheatstone Bridge
398
The utilization of strain gauges as transducers for converting mechanical strain into electrical signals is a common practice in various engineering applications. These strain gauges are frequently integrated into Wheatstone bridge circuits to accurately measure parameters such as force or pressure. Within this context, each element within the circuit exhibits a resistance that undergoes subtle variations when subjected to mechanical strain. The primary objective is to convert minuscule...
398