Low-dose CBCT protocols in implant dentistry: a systematic review
Laurits Kaaber1, Louise Hauge Matzen1, Lars Schropp1
1Department of Dentistry and Oral Health, Oral Radiology, Aarhus University, Aarhus, Denmark.
Low-dose cone-beam computed tomography (CBCT) protocols maintain objective image quality for dental implant therapy stages. Reducing radiation exposure is feasible without compromising crucial diagnostic accuracy.
Area of Science:
- Dentistry
- Radiology
- Medical Imaging
Background:
- Cone-beam computed tomography (CBCT) is integral to dental implant therapy.
- Optimizing CBCT protocols to reduce radiation exposure is a significant clinical goal.
- Evaluating low-dose protocols across implant planning, insertion, and peri-implantitis assessment is crucial.
Purpose of the Study:
- To review the current evidence on low-dose CBCT protocols in dental implant therapy.
- To assess the efficacy of reduced radiation protocols in implant planning, insertion, and follow-up.
- To determine the impact of low-dose CBCT on objective and subjective image quality.
Main Methods:
- Systematic literature search of MEDLINE/PubMed and Scopus databases.
- Inclusion of studies comparing low-dose CBCT protocols to reference standards for implant therapy.
- Data extraction and quality assessment of included studies.
Main Results:
- Sixteen studies were analyzed, identifying parameter reductions (kV, mA, voxel size, exposure time, trajectory) for low-dose protocols.
- Low-dose CBCT demonstrated comparable objective measurements to higher-dose protocols in all implant therapy stages.
- Bone measurements using CBCT were consistent with direct measurements, irrespective of the protocol.
- Subjective image quality may be adversely affected, while objective assessments remain largely unaffected.
Conclusions:
- Low-dose CBCT protocols do not compromise objective image quality in any stage of implant therapy.
- Parameter reductions, excluding kV, appear feasible for developing effective low-dose CBCT protocols.
- Further clinical research is warranted to confirm these findings and guide responsible radiation use in patient care.