Awareness for artifacts in fluorescence microscopy of β-TCP
Marco Waldmann1, Marc Bohner2, Anna Baghnavi3
1G.E.R.N. Tissue Replacement, Regeneration & Neogenesis, Department of Orthopedics and Trauma Surgery, Faculty of Medicine, Medical Center-Albert-Ludwigs-University of Freiburg, Albert-Ludwigs- University of Freiburg, Hugstetter Straße 55, 79106, Freiburg, Germany. mwaldmann1@gmx.de.
New artifacts in beta-tricalcium phosphate (β-TCP) ceramics can mimic cells under fluorescence analysis. Technovit fixation and mechanical cleaning effectively reduce these artifacts, preventing false positive results in cell imaging.
Area of Science:
- Biomaterials Science
- Cell Biology
- Microscopy Techniques
Background:
- Fluorescence analysis of beta-tricalcium phosphate (β-TCP) ceramics is a common method for characterizing cell presence.
- Previously undescribed artifacts resembling cells have been observed, potentially leading to misinterpretation in fluorescence imaging.
Purpose of the Study:
- To identify and characterize novel artifacts observed during fluorescence analysis of β-TCP ceramics.
- To evaluate methods for reducing these artifacts to prevent false positive cell identification.
Main Methods:
- Investigated artifacts in β-TCP ceramics using fluorescence microscopy.
- Applied prolonged ultrasound washing and Technovit 9100 fixation as artifact reduction techniques.
- Assessed the efficacy of mechanical cleaning in conjunction with Technovit fixation.
Main Results:
- Undescribed artifacts with cell-like shapes and fluorescence properties were identified.
- Untreated β-TCP ceramics showed no reduction in artifacts regardless of further treatment.
- Technovit 9100 fixation significantly reduced artifacts, with further reduction achieved through mechanical cleaning.
Conclusions:
- Artifacts in β-TCP ceramics can be mistaken for cells, necessitating careful evaluation.
- Technovit fixation, especially when combined with mechanical cleaning, is an effective method for reducing these artifacts.
- Researchers should be aware of these artifacts and employ artifact reduction strategies to ensure accurate cell detection in β-TCP ceramic studies.