Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
Evaluation of a method to quantify posture and scapula position using biplanar radiography
S Bousigues1, L Gajny1, W Skalli1
1Arts et Métiers, Institut de Biomécanique Humaine Georges Charpak, Paris, France.
This study introduces a new method for measuring scapular position and orientation using biplanar radiography. The technique demonstrates high accuracy and reproducibility, enabling better detection of pathological conditions.
Area of Science:
- Orthopedics
- Medical Imaging
- Biomechanical Engineering
Background:
- Quantifying scapular position and orientation in upright patients is crucial.
- Existing biplanar radiography methods struggle with consistent patient positioning for scapular and spinal imaging.
Purpose of the Study:
- To develop and validate a novel method for assessing scapular kinematics using upright biplanar radiographs.
- To establish reliable clinical parameters for scapular position and orientation.
Main Methods:
- Acquired two sets of biplanar radiographs for 3D reconstruction of the spine/ribcage and scapula.
- Performed two-stage scapular alignment (coarse and adjusted) based on anatomical landmarks.
- Computed clinical parameters: protraction, internal rotation, tilt, and upward rotation.
- Assessed reproducibility using in vivo data and accuracy using cadaveric CT scans.
Main Results:
- Mean error for clinical parameters was under 2°.
- 95% confidence intervals for reproducibility ranged from 2.5° to 5.3°.
- Confidence intervals were smaller than inter-participant variability.
Conclusions:
- The developed method offers high accuracy and reproducibility for assessing scapular kinematics.
- This technique shows potential for identifying distinct patterns in pathological populations.
- Enables precise 3D quantification of scapular position and orientation in upright individuals.