Screening the Implant-Augmented Breast with Digital Breast Tomosynthesis: Is Tomosynthesis Necessary for
Derek L Nguyen1, Lars J Grimm1, Jeffrey S Nelson1
1Department of Radiology, Duke University School of Medicine, Durham, NCUSA.
Objective:
To determine cancer visualization utility and radiation dose for non-implant-displaced (ID) views using standard protocol with digital breast tomosynthesis (DBT) vs alternative protocol with 2D only when screening women with implant augmentation.
Methods:
This retrospective cohort study identified women with implants who underwent screening DBT examinations that had abnormal findings from July 28, 2014, to December 31, 2021. Three fellowship-trained breast radiologists independently reviewed examinations retrospectively to determine if the initially identified abnormalities could be visualized on standard protocol (DBT with synthesized 2D (S2D) for ID and non-ID views) and alternate protocol (DBT with S2D for ID and only the S2D images for non-ID views). Estimated exam average glandular dose (AGD) and associations between cancer visualization with patient and implant characteristics for both protocols were evaluated.
Results:
The study included 195 patients (mean age 55 years ± 10) with 223 abnormal findings. Subsequent biopsy was performed for 86 abnormalities: 59 (69%) benign, 8 (9%) high risk, and 19 (22%) malignant. There was no significant difference in malignancy visualization rate between standard (19/223, 8.5%) and alternate (18/223, 8.1%) protocols (P = .92), but inclusion of the DBT for non-ID views found one additional malignancy. Total examination AGD using standard protocol (21.9 mGy ± 5.0) was significantly higher than it would be for estimated alternate protocol (12.6 mGy ± 5.0, P <.001). This remained true when stratified by breast thickness: 6.0-7.9 cm, 8.0-9.9 cm, >10.0 cm (all P <.001).
Conclusion:
The inclusion of DBT for non-ID views did not significantly increase the cancer visualization rate but did significantly increase overall examination AGD.