Experimental measurement of local noise power spectrum (NPS) in photon counting detector-CT (PCD-CT) using a single
Chengzhu Zhang1, Ke Li1,2, Ran Zhang1
1Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
Background:
Accurate noise power spectra (NPS) measurement in clinical X-ray CT exams is challenging due to the need for repeated scans, which expose patients to high radiation risks. A reliable method for single CT acquisition NPS estimation is thus highly desirable.
Purpose:
To develop a method for estimating local NPS from a single photon counting detector-CT (PCD-CT) acquisition.
Methods:
A novel nearly statistical bias-free estimator was constructed from the raw counts data of PCD-CT scan to estimate the variance of sinogram projection data. An analytical algorithm is employed to reconstruct point-wise covariance between any two image pixel/voxel locations and . A Fourier transform is applied to obtain the desired point-wise NPS for any chosen location . The method was validated using experimental data acquired from a benchtop PCD-CT system with various physical phantoms, and the results were compared with the conventional local NPS measurement method using repeated scans and statistical ensemble averaging.
Results:
The experimental results demonstrate that (1) the proposed method can achieve pointwise/local NPS measurement for a region of interest (ROI) located at any chosen position, accurately characterizing the NPS with spatial structures resulting from image content heterogeneity; (2) the local NPS measured using the proposed method show a higher precision in the measured NPS compared to the conventional measurement method; (3) spatial averaging of the local NPS yields the conventional NPS for a given local ROI.
Conclusion:
A new method was developed to enable local NPS from a single PCD-CT acquisition.