Measurement uncertainty in clinical chemistry: ISO 20914 versus nordtest or intermediate precision versus bias
Nilhan Nurlu1, Abdulkadir Cat1, Kamil Taha Ucar2
1Istanbul Gaziosmanpasa Training and Research Hospital, Department of Medical Biochemistry, Health Science University, Istanbul, Turkey.
Aim:
Measuring uncertainty (MU) is crucial to ensure the accuracy and precision of laboratory results. This study compares the ISO 20914 and Nordtest guidelines to analyze the MU values for 20 clinical chemistry analytes over six months.
Methods:
The researchers calculated MU components, including within-laboratory reproducibility (Rw), laboratory analytical performance bias (u(bias)), and combined standard uncertainty (uc), based on internal quality control and external quality assessment data. The final expanded uncertainty (U) values were determined by multiplying the combined uncertainty with a coverage factor (k = 2 for 95% Confidence Interval), following each guideline's respective procedures. Clinical chemistry analytes were analyzed on Roche Cobas 6000 c501 auto analyzer (Roche Diagnostics, Mannheim, Germany) and manufacturer's kits were used analysis.
Results:
The results show that 11 out of 20 clinical chemistry analytes met the targeted maximum allowable measurement uncertainty (MAU) values when calculated according to ISO 20914 guideline. Also, 11 out of 20 clinical chemistry analytes' MU values met the MAU values with the Nordtest guideline's recommended calculations. However, some tests met the MAU in the ISO 20914 approach but not in the Nordtest guideline, and vice versa.
Conclusions:
The study found that intermediate precision (uRw) in the ISO 20914 approach and performance bias (u(bias)) in the Nordtest approach significantly impacted MU values. The research highlights the importance of standardization in MU calculation approaches across clinical laboratories. These findings have implications for patient care and clinical decision-making, emphasizing the importance of selecting appropriate laboratory guidelines for routine use.
Insights
This study compared ISO 20914 and Nordtest guidelines for measuring uncertainty (MU) in clinical chemistry. Both guidelines showed 11 of 20 analytes met maximum allowable uncertainty, highlighting the need for standardized MU calculation approaches.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Measurement Science
Background:
- Accurate measurement uncertainty (MU) is vital for reliable laboratory results.
- Standardized guidelines are essential for consistent MU calculations in clinical settings.
Purpose of the Study:
- To compare measurement uncertainty (MU) values derived from ISO 20914 and Nordtest guidelines.
- To assess the impact of different MU calculation approaches on 20 clinical chemistry analytes.
Main Methods:
- MU components (reproducibility, bias) were calculated using internal quality control and external quality assessment data.
- Expanded uncertainty (U) was determined using coverage factors (k=2) per ISO 20914 and Nordtest procedures.
- Analyte analysis was performed on a Roche Cobas 6000 c501 autoanalyzer.
Main Results:
- 11 out of 20 analytes met maximum allowable measurement uncertainty (MAU) under both ISO 20914 and Nordtest guidelines.
- Discrepancies were observed, with some analytes meeting MAU by one guideline but not the other.
Conclusions:
- Intermediate precision (uRw) and performance bias (u(bias)) significantly influenced MU values.
- Standardization of MU calculation methods is crucial for clinical laboratories.
- Selecting appropriate guidelines impacts patient care and clinical decision-making.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Uncertainty: Overview
Data Validation
Key parameters for method validation include:
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Uncertainty in Measurement: Reading Instruments
Accuracy and Precision