Measurement uncertainty in clinical chemistry: ISO 20914 versus nordtest or intermediate precision versus bias

Nilhan Nurlu1, Abdulkadir Cat1, Kamil Taha Ucar2

  • 1Istanbul Gaziosmanpasa Training and Research Hospital, Department of Medical Biochemistry, Health Science University, Istanbul, Turkey.

Abstract

Insights

This study compared ISO 20914 and Nordtest guidelines for measuring uncertainty (MU) in clinical chemistry. Both guidelines showed 11 of 20 analytes met maximum allowable uncertainty, highlighting the need for standardized MU calculation approaches.

Area of Science:

  • Clinical Chemistry
  • Laboratory Medicine
  • Measurement Science

Background:

  • Accurate measurement uncertainty (MU) is vital for reliable laboratory results.
  • Standardized guidelines are essential for consistent MU calculations in clinical settings.

Purpose of the Study:

  • To compare measurement uncertainty (MU) values derived from ISO 20914 and Nordtest guidelines.
  • To assess the impact of different MU calculation approaches on 20 clinical chemistry analytes.

Main Methods:

  • MU components (reproducibility, bias) were calculated using internal quality control and external quality assessment data.
  • Expanded uncertainty (U) was determined using coverage factors (k=2) per ISO 20914 and Nordtest procedures.
  • Analyte analysis was performed on a Roche Cobas 6000 c501 autoanalyzer.

Main Results:

  • 11 out of 20 analytes met maximum allowable measurement uncertainty (MAU) under both ISO 20914 and Nordtest guidelines.
  • Discrepancies were observed, with some analytes meeting MAU by one guideline but not the other.

Conclusions:

  • Intermediate precision (uRw) and performance bias (u(bias)) significantly influenced MU values.
  • Standardization of MU calculation methods is crucial for clinical laboratories.
  • Selecting appropriate guidelines impacts patient care and clinical decision-making.

Related Concept Videos

Uncertainty in Measurement: Accuracy and Precision03:37

Uncertainty in Measurement: Accuracy and Precision

Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value. 
73.7K
Uncertainty: Overview00:59

Uncertainty: Overview

In analytical chemistry, we often perform repetitive measurements to detect and minimize inaccuracies caused by both determinate and indeterminate errors. Despite the cares we take, the presence of random errors means that repeated measurements almost never have exactly the same magnitude. The collective difference between these measurements - observed values - and the estimated or expected value is called uncertainty. Uncertainty is conventionally written after the estimated or expected value.
551
Data Validation01:15

Data Validation

Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
161
Contaminants and Errors01:16

Contaminants and Errors

Effective sample preparation is crucial for accurate and reliable laboratory analysis. During this process, two significant sources of error can arise: concentration bias from improper sample splitting and contamination caused by methods used to reduce particle size, such as grinding or homogenization. Identifying and minimizing these potential errors is crucial to ensuring the validity of the analysis.
Another key consideration is determining the appropriate number of samples required to...
89
Uncertainty in Measurement: Reading Instruments02:46

Uncertainty in Measurement: Reading Instruments

Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
38.1K
Accuracy and Precision01:52

Accuracy and Precision

Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.  Highly accurate...
8.8K