MR Safety of Inductively Coupled and Conventional Intraoral Coils
Agazi Samuel Tesfai1, Simon Reiss, Thomas Lottner
1From the Division of Medical Physics, Department of Diagnostic and Interventional Radiology, University Medical Center Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Purpose:
Intraoral coils (IOCs) in magnetic resonance imaging (MRI) significantly improve the signal-to-noise ratio compared with conventional extraoral coils. To assess the safety of IOCs, we propose a 2-step procedure to evaluate radiofrequency-induced heating of IOCs and compare maximum temperature increases in 3 different types of IOCs.
Methods:
The 2-step safety assessment consists of electric field measurements and simulations to identify local hotspots followed by temperature measurements during MRI. With this method, 3 different coil types (inductively coupled IFC, transmit/receive tLoop, and receive-only tLoopRx) were tested at 1.5 T and 3 T for both tuned and detuned coil states. High SAR and regular MRI protocols were applied for 2 coil positions.
Results:
The measured E field maps display distinct hotspots for all tuned IOCs, which were reduced by at least 40-fold when the IOCs were detuned. Maximum temperature rise was higher when the coils were positioned at the periphery of the phantom with the coil planes parallel to B 0 . When neither active nor passive detuning was applied, maximum temperature increase of ΔT = 1.3/0.5/1.8 K was found for IFC/tLoop/tLoopRx coils. Hotspots detected by E field measurements, and simulations were consistent. In the simulations, the results were different for homogeneous phantoms compared with full anatomical models. The 2-step test procedure is applicable to different coil types.
Conclusions:
The results indicate that a risk for radiofrequency-induced heating exists for tuned IOCs, so that adequate detuning circuits need to be integrated in the coils to ensure safe operation.
More Related Videos
05:26Author Spotlight: Low-Cost Electroencephalographic Recording System Combined with a Millimeter-Sized Coil to Transcranially Stimulate the Mouse Brain In Vivo
Published on: May 26, 2023
11:27Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
Related Concept Videos
Induction
A...
Inductors
Mutual Inductance
When two circuits carrying time-varying currents are close to one another, the magnetic flux through each circuit varies because of the changing current in the other circuit. Consequently, an emf is induced in each circuit by the changing current in the other. Therefore, this type of emf is called...
Induced Electric Fields: Applications
Charging Conductors By Induction
Generally, conductors like metals do not allow any excess charge to be present on them. Any excess charge added to metals easily flows away, for example, when a metal is placed on the Earth. This process is called earthing.
However, conductors can be charged by a process called induction. For example, consider charging a...
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...