Abdominal image quality and dose reduction with energy-integrating or photon-counting detectors dual-source CT: A
Joël Greffier1, Djamel Dabli1, Sebastian Faby2
1IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
Purpose:
The purpose of this study was to assess image-quality and dose reduction potential using a photon-counting computed tomography (PCCT) system by comparison with two different dual-source CT (DSCT) systems using two phantoms.
Materials And Methods:
Acquisitions on phantoms were performed using two DSCT systems (DSCT1 [Somatom Force] and DSCT2 [Somatom Pro.Pulse]) and one PCCT system (Naeotom Alpha) at four dose levels (13/6/3.4/1.8 mGy). Noise power spectrum (NPS) and task-based transfer function (TTF) were computed to assess noise magnitude and noise texture and spatial resolution (f50), respectively. Detectability indexes (d') were computed to model the detection of abdominal lesions: one unenhanced high-contrast task, one contrast-enhanced high-contrast task and one unenhanced low-contrast task. Image quality was subjectively assessed on an anthropomorphic phantom by two radiologists.
Results:
For all dose levels, noise magnitude values were lower with PCCT than with DSCTs. For all CT systems, similar noise texture values were found at 13 and 6 mGy, but the greatest noise texture values were found for DSCT2 and the lowest for PCCT at 3.4 and 1.8 mGy. For high-contrast inserts, similar or lower f50 values were found with PCCT than with DSCT1 and the opposite pattern was found for the low-contrast insert. For the three simulated lesions, d' values were greater with PCCT than with DSCTs. Abdominal images were rated satisfactory for clinical use by the radiologists for all dose levels with PCCT and for 13 and 6 mGy with DSCTs.
Conclusion:
By comparison with DSCTs, PCCT reduces image-noise and improves detectability of simulated abdominal lesions without altering the spatial resolution and image texture. Image-quality obtained with PCCT seem to indicate greater potential for dose optimization than those obtained with DSCTs.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...