Evaluation of Hemodynamic Alterations after Flow Diverter Placement using the AneurysmFlowTM tool
V Dang Luu1, T Xuan Bach1, N Huu An1
1Radiology Center, Bach Mai Hospital, Hanoi, Vietnam.
This study evaluated AneurysmFlow software for assessing flow changes after flow-diverting treatment in aneurysms. Quantitative flow changes did not correlate with aneurysm occlusion status at 6 months.
Area of Science:
- Medical Imaging
- Interventional Radiology
- Biomedical Engineering
Background:
- AneurysmFlow, a novel optical flow-based tool from DSA, requires further validation.
- Flow-diverting stent (FDS) treatment is a key intervention for aneurysms.
Purpose of the Study:
- To assess the significance of flow velocity changes and Mean Aneurysm Flow Amplitude (MAFA) ratio in predicting outcomes after FDS treatment.
- To evaluate the utility of AneurysmFlow software in post-treatment aneurysm assessment.
Main Methods:
- Retrospective analysis of 41 patients (42 aneurysms) treated with FDS and measured using AneurysmFlow.
- Analysis of flow velocity changes and MAFA ratio in relation to aneurysm characteristics and occlusion status at 6-month follow-up.
Main Results:
- AneurysmFlow achieved a 90.5% success rate in 38 aneurysms.
- A decrease in flow velocity was observed in 78.9% of cases post-stent deployment.
- No significant association was found between flow changes, MAFA ratio, and aneurysm size or occlusion status at 6 months.
Conclusions:
- Quantitative flow changes measured by AneurysmFlow software did not correlate with aneurysm occlusion status at 6 months post-FDS treatment.
- Larger studies with longer follow-up are needed to further investigate these findings and the role of AneurysmFlow.
More Related Videos
12:00Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
11:16In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
Related Concept Videos
Blood Flow
Steady Flow of a Fluid Stream
During this process, the momentum of the fluid within the control volume remains constant over the time interval dt. By applying the...
Pipe Flowrate Measurement
The orifice meter is a simple,...
General External Flow Characteristics
Rapidly Varying Flow
Applications of Integration to Find Blood Flow