Biomechanical evaluation of reinsertion and revision screws in the subaxial cervical vertebrae
Wei-Xin Dong1, Yong Hu2, Ou-Jie Lai1
1Department of Spinal Surgery, Ningbo No.6 Hospital, 1059 East Zhongshan Road, Ningbo, Zhejiang, 315040, People's Republic of China.
Background:
This study aimed to evaluate the biomechanical effects of reinserted or revised subaxial cervical vertebral screws.
Methods:
The first part aimed to gauge the maximum insertional torque (MIT) of 30 subaxial cervical vertebrae outfitted with 4.0-mm titanium screws. A reinsertion group was created wherein a screw was wholly removed and replaced along the same trajectory to test its maximum pullout strength (MPOS). A control group was also implemented. The second part involved implanting 4.0-mm titanium screws into 20 subaxial cervical vertebrae, testing them to failure, and then reinserting 4.5-mm revision screws along the same path to determine and compare the MIT and MPOS between the test and revision groups.
Results:
Part I findings: No significant difference was observed in the initial insertion's maximum insertion torque (MIT) and maximum pull-out strength (MPOS) between the control and reinsertion groups. However, the MIT of the reinsertion group was substantially decreased compared to the first insertion. Moderate to high correlations were observed between the MIT and MPOS in both groups, as well as between the MIT of the first and second screw in the reinsertion group. Part II, the MIT and MPOS of the screw in the test group showed a strong correlation, while a modest correlation was observed for the revision screw used in failed cervical vertebrae screw. Additionally, the MPOS of the screw in the test group was significantly higher than that of the revision screw group.
Conclusion:
This study suggests that reinsertion of subaxial cervical vertebrae screws along the same trajectory is a viable option that does not significantly affect fixation stability. However, the use of 4.5-mm revision screws is inadequate for failed fixation cases with 4.0-mm cervical vertebral screws.
More Related Videos
Related Concept Videos
Bone Remodeling
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...
Bone Remodeling and Repair
Normal Strain under Axial Loading
Eccentric Axial Loading in a Plane of Symmetry
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...