Low-contrast detectability of photon-counting-detector CT at different scan modes and image types in comparison with
Mingdong Fan1, Zhongxing Zhou1, Jarod Wellinghoff1
1Mayo Clinic, Department of Radiology, Rochester, Minnesota, United States.
Purpose:
We aim to compare the low-contrast detectability of a clinical whole-body photon-counting-detector (PCD)-CT at different scan modes and image types with an energy-integrating-detector (EID)-CT.
Approach:
We used a channelized Hotelling observer (CHO) previously optimized for quality control purposes. An American College of Radiology CT accreditation phantom was scanned on both PCD-CT and EID-CT with 10 phantom positionings. For PCD-CT, images were generated using two scan modes, standard resolution (SR) and ultra-high-resolution (UHR); two image types, virtual monochromatic images at 70 keV and low-energy threshold (T3D); both filtered-back-projection (FBP) and iterative reconstruction (IR) reconstruction methods; and three reconstruction kernels. For each positioning, three repeated scans were acquired for each scan mode, image type, and CTDIvol of 6, 12, and 24 mGy. For EID-CT, images acquired from scans (10 positionings × 3 repeats × 3 doses) were reconstructed using the closest counterpart FBP and IR kernels. CHO was applied to calculate the index of detectability () on both scanners.
Results:
With the smooth Br44 kernel, the of UHR was mostly comparable with that of the SR mode (difference: -11.4% to 8.3%, to 0.956), and the T3D images had a higher (difference: 0.7% to 25.6%) than 70 keV images on PCD-CT. Compared with the EID-CT, UHR-T3D of PCD-CT had non-inferior (difference: -2.7% to 12.9%) with IR and non-superior (difference: 0.8% to 11.2%) with FBP using the Br44 kernel. PCD-CT produced higher than EID-CT by 61.8% to 247.1% with the sharper reconstruction kernels.
Conclusions:
The comparison between PCD-CT and EID-CT was significantly influenced by the reconstruction method and kernel. With a smooth kernel that is typically used in low-contrast detection tasks, the PCD-CT demonstrated low-contrast detectability that was comparable to EID-CT with IR and showed no superiority when using FBP. With the use of sharper kernels, the PCD-CT significantly outperformed EID-CT in low-contrast detectability.
More Related Videos
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...