Related Experiment Video
Updated: Jun 25, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Real-time gated proton therapy with a reduced source to imager distance: Commissioning and quality assurance
Hong Qi Tan1, Calvin Wei Yang Koh2, Kah Seng Lew2
1Division of Radiation Oncology, National Cancer Centre Singapore, Singapore; Oncology Academic Clinical Programme, Duke-NUS Medical School, Singapore.
Introduction:
Real-time gated proton therapy (RGPT) is a motion management technique unique to the Hitachi particle therapy system. It uses pulsed fluoroscopy to track an implanted fiducial marker. There are currently no published guidelines on how to conduct the commissioning and quality assurance. In this work we reported on our centre's commissioning workflow and our daily and monthly QA procedures.
Methods:
Six commissioning measurements were designed for RGPT. The measurements include imaging qualities, fluoroscopic exposures, RGPT marker tracking accuracy, temporal gating latency, fiducial marker tracking fidelity and an end-to-end proton dosimetry measurement. Daily QA consists of one measurement on marker localization accuracy. Four months daily QA trends are presented. Monthly QA consists of three measurementson the gating latency, fluoroscopy imaging quality and dosimetry verification of gating operation with RGPT.
Results:
The RGPT was successfully commissioned in our centre. The air kerma rates were within 15 % from specifications and the marker tracking accuracies were within 0.245 mm. The gating latencies for turning the proton beam on and off were 119.5 and 50.0 ms respectively. The 0.4x10.0 mm2 Gold AnchorTM gave the best tracking results with visibility up to 30 g/cm2. Gamma analysis showed that dose distribution of a moving and static detectors had a passing rate of more than 95 % at 3 %/3mm. The daily marker localization QA results were all less than 0.2 mm.
Conclusion:
This work could serve as a good reference for other upcoming Hitachi particle therapy centres who are interested to use RGPT as their motion management solution.