Temperature Changes in the Pulp Chamber During Orthodontic Bonding Using a High-Power Light-Emitting Diode Device
Alia Alsafadi1, Nada Rajeh1, Mohammad Y Hajeer1
1Department of Orthodontics, Faculty of Dentistry, University of Damascus, Damascus, SYR.
Objective:
This study used a high-power light-emitting diode (LED) device to evaluate the effects of two exposure times and intensities on pulp chamber temperature and cooling time during bracket bonding.
Materials And Methods:
Sixty upper premolars were used in the sample in this study. These premolars were split into two main groups based on the exposure time and intensity: the first group employed a traditional curing mode (TCG) for 20 seconds with an intensity of 1200 mw/cm2, whereas the second group had a quick curing mode (QCG) for 3 seconds with an intensity of 2500 mw/cm2. The pulp chamber's temperature variations and cooling times were recorded using a thermal imaging camera. The Mann-Whitney U test was used to find differences between the two-group comparison of the pulp chamber's temperature and cooling time.
Results:
The two groups had statistically significant differences regarding the temperature increase in the pulp chamber and cooling time (p > 0.001). The mean temperature increase in the traditional curing group was 3.52°C, which is greater than that in the quick curing group (i.e., a mean value of 1.28°C). The mean cooling time in the traditional curing group was 38.83 seconds, which is greater than that in the quick curing group (9.97 seconds).
Conclusions:
Reducing the exposure time to 3 seconds and increasing the intensity to 2500 mw/cm2 is considered safer for the pulp chamber during and after the curing process.