Multicentric characterization of organ-based tube current modulation in head computed tomography: A dosimetric and
Séléna Pondard1, Corentin Desport1, Mélodie Munier2
1Alara Expertise, 7 Allée de l'Europe, 67960 Entzheim, France.
Purpose:
To evaluate the efficiency of organ-based tube current modulation (OBTCM) in head Computed Tomography (CT) for different radiology departments and manufacturers.
Materials And Methods:
Five CT scanners from four radiology departments were evaluated in this study. All scans were performed using a standard and a routine head protocol. A scintillating fiber optic detector was placed directly on the gantry to measure the tube exit kerma. Image quality was quantified on a 16-cm HEAD phantom by measuring the signal-to-noise ratio (SNR) and the standard deviation of the Hounsfield units (HU) of circular regions of interest placed in the phantom. The Noise Power Spectrum (NPS) was also studied. Measured values were compared on images with and without OBTCM.
Results:
The reduction rates in tube exit kerma, on the anterior part, vary between 11 % and 74 % depending on the CT scanner and the protocol used. The tube exit kerma on the posterior part remains unchanged in GE and Canon CT scanners. On the contrary, the tube exit kerma to the posterior part increases by up to 39 % in Siemens CT scanner. Image noise and SNR increase by up to 10 % in the five CT scanners. Nonetheless, the differences in noise and SNR are statistically significant (p-value < 0.05).The analysis of the NPS indicates that the noise texture remains unchanged.
Conclusion:
OBTCM reduces the tube exit kerma to the anterior part of the gantry without reducing substantially image quality for head protocols.
More Related Videos
07:52Author Spotlight: Advancing Human Brain Modulation – Optimized Protocols for Transcranial Ultrasound Stimulation Experiments
Published on: June 28, 2024
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...