Clinical audit of the image quality and customised contrast volume using P3T contrast injection software versus
N Jayamani1, S Pothiawala2, H B Ong1
1Department of Radiography, Singapore General Hospital, Singapore.
Introduction:
The implications of shorter scan time and lower tube voltage in the dual-source CT coronary angiography (CTCA) scan protocol necessitate the adaptation of contrast media (CM) injection parameters. This audit evaluates the coronary arteries' vascular attenuation and image quality by comparing the personalised patient protocol technology (P3T) contrast injection software with standard injection protocol. The secondary aim is to determine the relationship between CM volume and the patient's weight.
Methodology:
A Siemens Somatom Definition Force CT Unit was used to scan 30 sets of patients between August 2020 and October 2020. Patients were selected retrospectively and separated into Standard Injection and P3T injection protocols. An experienced radiologist blinded to the groups reviewed the coronary vessels' contrast enhancement and image quality.
Results:
Overall, the mean HU of all the main coronary artery vessels obtained from P3T injection software reached above 350 HU and was diagnostically sufficient. The mean attenuation at the proximal region of RCA in the 80-99 kg weight category was significantly higher in the P3T injection software than the standard injection protocol (p < 0.001). The CM volume proposed by P3T injection software for 40-59 kg was approximately 57 ± 5 mls, while 75 ml was used for the standard injection protocol.
Conclusion:
P3T injection software in CTCA resulted in an adequate diagnostic attenuation of coronary arteries (>350HU) in all weight groups, most effectively in the higher weight group, while maintaining diagnostic image quality. Further, the P3T software reduces CM volumes in lower-weight patients.
Implications:
P3T software enables reducing CM volume in lower-weight patients while improving vascular enhancement in CTCA scans in higher-weight patients.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...