Related Experiment Video
Updated: Jun 24, 2025

An All-in-one Sample Holder for Macromolecular X-ray Crystallography with Minimal Background Scattering
Published on: July 6, 2019
Conceptual design of a macromolecular diffractometer for the Jülich high brilliance source
Z Ma1,2, K Lieutenant1, J Voigt1
1Jülich Centre for Neutron Science, Forschungszentrum Jülich, 52425 Jülich, Germany.
Abstract:
In this work, a concept for a neutron diffractometer for high-resolution macromolecular structures has been developed within the Jülich High Brilliance Neutron Source (HBS) project. The SELENE optics are adapted to the requirements of the instrument to achieve a tunable low background neutron beam at mm2 scale sample area. With the optimized guide geometry, a low background neutron beam can be achieved at the small sample area with tunable divergence and size. For the 1 × 1 mm2 sample, a flux of 1.10 × 107 n/s/cm2 for 0.38° divergence is calculated in the 2-4 Å wavelength range, which is about 84.6% of the flux at MaNDi of the high-power spallation source SNS at ORNL. Virtual neutron scattering experiments have been performed to demonstrate the instrument's capabilities for studies of mm scale samples with large unit cells. Results of Vitesse simulations indicate that unit cell sizes of up to 200 Å are possible to be resolved with the proposed instrument.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
07:42On-Chip Crystallization and Large-Scale Serial Diffraction at Room Temperature
Published on: March 11, 2022