Prostate size, source configuration, and dosimetry dynamics of stranded 125I seed implants
Ron S Sloboda1, Derek Liu1, Nick Chng2
1Cross Cancer Institute, Edmonton, Alberta, Canada.
Seed movement after prostate brachytherapy is mostly random, influenced by edema. While prostate size changes are complex to infer from seed movement, dosimetry metrics like V100 and D90 remain relatively stable post-implant.
Area of Science:
- Radiation Oncology
- Medical Imaging
- Urology
Background:
- Transperineal implantation of stranded 125I seeds is a common treatment for prostate cancer.
- Accurate dosimetry is crucial for treatment efficacy and minimizing side effects.
- Understanding prostate size changes and seed migration post-implantation is essential for dose assessment.
Purpose of the Study:
- To quantify prostate size changes and seed movement following 125I brachytherapy.
- To assess the impact of these changes on prostate dosimetry over time.
- To evaluate the reliability of seed movement for inferring prostate dimensional changes.
Main Methods:
- CT and MR imaging (T2, balanced steady-state free precession) were performed on 20 patients at days 0, 3, 10, and 30 post-implantation.
- Automated image registration was used to align sequential scans.
- Prostate contours and seed positions were analyzed to calculate volume, dimensions, seed movement, and dose metrics (V200, V150, V100, D90).
Main Results:
- Prostate dimensions in the anterior-posterior and left-right directions showed modest reductions consistent with contour measurements.
- Prostate volume and superior-inferior dimension changes were overestimated when inferred from seed movements.
- Average seed movement was 4.8 ± 3.0 mm, primarily attributed to prostate edema resolution.
- By day 30, cohort-averaged dose metrics showed ratios relative to day 0 of 1.67 for V200, 1.33 for V150, 1.02 for V100, and 1.08 for D90.
Conclusions:
- Prostate size reduction in the implant direction cannot be reliably determined from stranded seed movements.
- Residual seed movements, excluding edema resolution, are largely random.
- Prostate V100 and D90 show modest changes by day 30, while V150 and V200 experience substantial increases.