A simple method to overcome paraproteinemic interferences in chemistry and immunoassays
1Drs. Tribedi & Roy Diagnostic Laboratory, Kolkata, India.
Background:
Interferences on chemistry and immunoassay results due to paraproteinemia may lead to erroneous diagnoses and treatment. Such interferences are difficult to recognize and even more difficult to deal with. This report describes 1 such case where multiple measurands were affected and how the interferant was overcome.
Case Report:
Paraproteins present in an immunoglobulin (Ig)G-lambda multiple myeloma specimen interfered with results of total bilirubin, direct bilirubin, inorganic phosphate, iron, ferritin, and total thyroxine measured on 3 platforms: AU5800, Alinity ci, and cobas pure. Repeat testing upon dilution with normal saline or deproteinization by polyethylene glycol precipitation gave unsatisfactory results on some or all the affected measurands. Repeat testing after dilution of the interferant serum with a healthy serum corrected the anomalous results for all the affected measurands.
Conclusion:
Dilution of paraproteinemic serum with a healthy serum of known concentrations appears to be the most suitable method to negate the effects of paraproteinemic interferences.
Related Concept Videos
Immunoprecipitation
Chromatin Immunoprecipitation
Chromatin immunoprecipitation, also known as ChIP, is used to study protein-DNA or...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences