Investigation of intra-fractionated range guided adaptive proton therapy: I. On-line PET imaging and range
Dongxu Yang1, Xiaorong R Zhu2, Mingli Chen1
1Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX 75057, United States of America.
Abstract:
Objective.Develop a prototype on-line positron emission tomography (PET) scanner and evaluate its capability of on-line imaging and intra-fractionated proton-induced radioactivity range measurement.Approach.Each detector consists of 32 × 32 array of 2 × 2 × 30 mm3Lutetium-Yttrium Oxyorthosilicate scintillators with single-scintillator-end readout through a 20 × 20 array of 3 × 3 mm2Silicon Photomultipliers. The PET can be configurated with a full-ring of 20 detectors for conventional PET imaging or a partial-ring of 18 detectors for on-line imaging and range measurement. All detector-level readout and processing electronics are attached to the backside of the system gantry and their output signals are transferred to a field-programable-gate-array based system electronics and data acquisition that can be placed 2 m away from the gantry. The PET imaging performance and radioactivity range measurement capability were evaluated by both the offline study that placed a radioactive source with known intensity and distribution within a phantom and the online study that irradiated a phantom with proton beams under different radiation and imaging conditions.Main results.The PET has 32 cm diameter and 6.5 cm axial length field-of-view (FOV), ∼2.3-5.0 mm spatial resolution within FOV, 3% sensitivity at the FOV center, 18%-30% energy resolution, and ∼9 ns coincidence time resolution. The offline study shows the PET can determine the shift of distal falloff edge position of a known radioactivity distribution with the accuracy of 0.3 ± 0.3 mm even without attenuation and scatter corrections, and online study shows the PET can measure the shift of proton-induced positron radioactive range with the accuracy of 0.6 ± 0.3 mm from the data acquired with a short-acquisition (60 s) and low-dose (5 MU) proton radiation to a human head phantom.Significance.This study demonstrated the capability of intra-fractionated PET imaging and radioactivity range measurement and will enable the investigation on the feasibility of intra-fractionated, range-shift compensated adaptive proton therapy.
More Related Videos
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...