Related Experiment Video
Updated: Jun 24, 2025

Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
Impact of Extracorporeal Membrane Oxygenation Circuitry on Remdesivir
Jeffrey J Cies1,2,3, Wayne S Moore1, Jillian Deacon2
1The Center for Pediatric Pharmacotherapy LLC (JJC, WSM, AC), Pottstown, PA.
Objectives:
This study aimed to determine the oxygenator impact on alterations of remdesivir (RDV) in a contemporary neonatal/pediatric (1/4-inch) and adolescent/adult (3/8-inch) extracorporeal membrane -oxygenation (ECMO) circuit including the Quadrox-i oxygenator.
Methods:
One-quarter-inch and a 3/8-inch, simulated closed-loop ECMO circuits were prepared with a Quadrox-i pediatric and Quadrox-i adult oxygenator and blood primed. Additionally, 1/4-inch and 3/8-inch circuits were also prepared without an oxygenator in series. A 1-time dose of RDV was administered into the circuits and serial preoxygenator and postoxygenator concentrations were obtained at 0 to 5 minutes, and 1-, 2-, 3-, 4-, 5-, 6-, 8-, 12-, and 24-hour time points. The RDV was also maintained in a glass vial and samples were taken from the vial at the same time periods for control purposes to assess for spontaneous drug degradation.
Results:
For the 1/4-inch circuits with an oxygenator, there was a 35% to 60% RDV loss during the study period. For the 1/4-inch circuits without an oxygenator, there was a 5% to 20% RDV loss during the study period. For the 3/8-inch circuit with and without an oxygenator, there was a 60% to 70% RDV loss during the study period.
Conclusions:
There was RDV loss within the circuit during the study period and the RDV loss was more pronounced with the larger 3/8-inch circuit when compared with the 1/4-inch circuit. The impact of the -oxygenator on RDV loss appears to be variable and possibly dependent on the size of the circuit and -oxygenator. These preliminary data suggest RDV dosing may need to be adjusted for concern of drug loss via the ECMO circuit. Additional single- and multiple-dose studies are needed to validate these findings.
More Related Videos
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
07:41Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Mechanical Ventilation I: Indication and Settings