Silicone implant and fibrous capsule assessment based on water-fat-silicone chemical shift encoding-based species
Tabea Borde1, Jonathan Stelter1, Antonia Wiedemann1
1Department of Diagnostic and Interventional Radiology, School of Medicine & Health, Technical University of Munich, Munich, Germany.
Background:
With rising breast augmentations worldwide, there is an increasing clinical need for an early and accurate detection of implant complications.
Purpose:
To compare the quality of chemical shift encoding-based (CSE) water-fat-silicone separation compared to double inversion recovery (DIR) silicone-only imaging in breast magnetic resonance imaging (MRI).
Material And Methods:
This retrospective, single-center study included women with silicone implants subjected to 3-T MRI between January 2021 and March 2022. MRI included (i) two-dimensional silicone-only T2-weighted turbo spin echo DIR acquisition and (ii) three-dimensional CSE imaging based on multi-echo gradient-echo sequence enabling water-, fat-, and silicone-image separation. Images were evaluated and compared by three independent radiologists using a clinically established rating including differentiability of the silicone implant, visibility and contouring of the adjacent fibrous capsule, and accuracy of intralesional folds in a ranking of 1-5. The apparent contrast-to-noise (aCNR) was calculated.
Results:
In 71 women, the average quality of water-fat-silicone images from CSE imaging was assessed as "good" (assessment 4 ± 0.9). In 68 (96%) patients, CSE imaging achieved a concise delineation of the silicone implant and precise visualization of the fibrous capsule that was not distinguishable in DIR imaging. Implant ruptures were more easily detected in CSE imaging. The aCNR was higher in CSE compared to DIR imaging (18.43 ± 9.8 vs. 14.73 ± 2.5; P = 0.002).
Conclusion:
Intrinsically co-registered water-fat-silicone-separated CSE-based images enable a reliable assessment of silicone implants. The simultaneously improved differentiability of the implant and fibrous capsule may provide clinicians with a valuable tool for an accurate evaluation of implant integrity and early detection of potential complications.