Related Experiment Video
Updated: Jun 23, 2025

Ex Vivo Porcine Experimental Model for Studying and Teaching Lung Mechanics
Published on: April 19, 2024
Development and Safety Trial of the OstreaVent2™ Prototype for Mechanically Ventilated Adult Patients
Maria Esterlita T Villanueva-Uy1,2, Enrique M Ostrea1, Alexander P Paran3
1Institute of Child Health and Human Development, National Institutes of Health, University of the Philippines Manila, Manila, Philippines.
Background:
With the surge of COVID-19 infections, there were concerns about shortage of mechanical ventilator in several countries including the Philippines.
Objective:
To transform a locally made, low-cost, neonatal ventilator into a volume- and pressure-controlled, adult ventilator and to determine its safe use among ventilated, adult patients at the Philippine General Hospital.
Methods:
The modification of the neonatal ventilator (OstreaVent1) to the adult OstreaVent2 was based on the critical need for adult ventilators, in volume or pressure mode, in the Philippines due to the COVID-19 pandemic. The adult ventilator settings were calibrated and tested for two days to check for consistency and tolerance and then submitted to a third party for certification. Once certified, a safety trial of 10 stable adult patients on mechanical ventilator was conducted. The patients were placed on the OstreaVent2 for four hours while ventilator parameters, patient's vital signs, and arterial blood gases were monitored at baseline, during, and after placement on the OstreaVent2. A post-study chest radiograph was also done to rule out pulmonary complications, particularly atelectasis and pneumothorax.
Results:
The prototype OstreaVent2 received an FDA Certification for Medical Listing after passing its third-party certification. Ten patients (60% male) recruited in the study had a mean age of 39.1 ± 11.6 years. Half of the patients had a diagnosis of non-COVID-19 pneumonia. During the 4-hour study period, the patients while on the OstreaVent2, had stable ventilator settings and most of the variabilities were within the acceptable tolerances. Vital signs were stable and arterial blood gases were within normal limits. One patient developed alar flaring which was relieved by endotracheal tube suctioning. No patient was withdrawn from the study. One patient who was already transferred out of the ICU subsequently deteriorated and died three days after transfer to the stepdown unit from a non-ventilator related cause.
Conclusion:
The new OstreaVent2 is safe to use among adults who need ventilator support. Variabilities in the ventilator's performance were within acceptable tolerances. Clinical and blood gas measurements of the patients were stable while on the ventilator.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...