Background:
Computed tomography (CT)-body divergence limits the accuracy of electromagnetic navigation bronchoscopy (ENB) in peripheral lung lesions diagnosis. We developed intraprocedural CT-guided navigation with ventilatory strategy for atelectasis (ICNVA) ENB for patients with peripheral lung lesions.
Methods:
Retrospective observational study in which ten consecutive patients with pulmonary lesions (without bronchial direct connection) underwent ICNVA-ENB was conducted. During ICNVA-ENB, intraoperative CT data were used for ENB path planning, and a new ventilation strategy were employed to help maintain the pulmonary region in a static and inflation state which reduce CT to body divergence. We collected three sets of CT data: preENB CT, post-anesthesia intubation CT, and postENB CT. To evaluate the accuracy of ICNVA-ENB, we measured the distance between the ENB probe and the actual lesion location, but also recorded the results of rapid on-site evaluation (ROSE), and postoperative pathology. To evaluate the impact of CT-body divergence induced by atelectasis, we calculated the mutual position distance of target lesions in preENB CT, post-anesthesia intubation CT and postENB CT. Furthermore, ENB operation time and operative complications were recorded.
Results:
Our analysis revealed that the distance between the navigation probe with the actual location of lesion center was 4-10 (5.90±1.73) mm. The ROSE results were consistent with the postoperative pathological diagnosis in 9 out of 10 patients (90%). The ICNVA-ENB atelectasis CT-body divergence was smaller than traditional ENB (12.10±3.67 vs. 6.60±2.59 mm, P<0.01). The ENB operation time was 20-53 (29.30±10.14) minutes and one patient developed slight intrapulmonary hemorrhage.
Conclusions:
ICNVA-ENB can reduce the CT-body divergence and appears to be safe and accurate for patients with peripheral lung lesions.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...