Ultrasound Cyclo Plasty in Eyes with Glaucoma
Published on: January 26, 2018
Uveitis-Glaucoma-Hyphema Syndrome From Subtle Malposition of a Hydrus Microstent
Ashley Polski1, Theresa Long, Brian C Stagg
1Department of Ophthalmology, Moran Eye Center, University of Utah, Salt Lake City, UT.
None:
Uveitis-glaucoma-hyphema (UGH) syndrome is a rare condition characterized by intraocular inflammation, elevated intraocular pressure, and recurrent hyphema. Although UGH syndrome is typically caused by chafing between an intraocular lens and the iris or ciliary body, microinvasive glaucoma devices are becoming increasingly recognized as potential causes of recurrent intraocular inflammation. Herein, we discuss a case of UGH syndrome due to the subtle malposition of a Hydrus microstent that ultimately required surgical removal to manage. This represents the first case of Hydrus-related UGH syndrome to include high-quality anterior segment imaging, which led us to suspect Hydrus-iris contact and supported the decision to proceed with microstent explantation.