Related Experiment Video
Updated: Jun 23, 2025

Force System with Vertical V-Bends: A 3D In Vitro Assessment of Elastic and Rigid Rectangular Archwires
Published on: July 24, 2018
Effect of thermomechanical ageing on force transmission by orthodontic aligners made of different thermoformed
Tarek M Elshazly1, Christoph Bourauel1, Ahmed M Ismail1,2
1Oral Technology, Dental School, University Hospital Bonn, Bonn, Germany.
Objectives:
Investigating the impact of thermal and mechanical loading on the force generation of orthodontic aligners made from various thermoplastic materials and different compositions.
Materials And Methods:
Five distinct materials were utilized including, three multi-layer (Zendura FLX, Zendura VIVA, CA Pro) and two single-layer (Zendura A and Duran). A total of 50 thermoformed aligners (n = 10) underwent a 48-hour ageing protocol, which involved mechanical loading resulting from a 0.2 mm facial malalignment of the upper right central incisor (Tooth 11) and thermal ageing through storage in warm distilled water at 37°C. The force exerted on Tooth 11 of a resin model was measured both before and after ageing using pressure-sensitive films and a biomechanical setup.
Results:
Before ageing, pressure-sensitive films recorded normal contact forces ranging from 83.1 to 149.7 N, while the biomechanical setup measured resultant forces ranging from 0.1 to 0.5 N, with lingual forces exceeding facial forces. Multi-layer materials exhibited lower force magnitudes compared to single-layer materials. After ageing, a significant reduction in force was observed, with some materials experiencing up to a 50% decrease. Notably, multi-layer materials, especially Zendura VIVA, exhibited lower force decay.
Conclusions:
The force generated by aligners is influenced by both the aligner material and the direction of movement. Multi-layer materials exhibit superior performance compared to single-layer materials, primarily because of their lower initial force, which enhances patient comfort, and their capability to maintain consistent force application even after undergoing ageing.
Related Concept Videos
Thermal Strain
Temperature Dependent Deformation
Thermal Stress
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in...
Bone Disorders
Bone deposition is also affected by the levels of sex hormones like estrogen and testosterone that promote osteoblast activity and bone matrix synthesis. When the level of these hormones decreases due to aging, it causes a reduction in bone deposition. As a result, bone resorption by osteoclasts...
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity