Technical note: Quantifying radionuclide residues in hospital wastewater: A case study on [131I]I and
Dennis Kupitz1, Eric Einspänner1, Heiko Wissel1
1Department for Radiology and Nuclear Medicine, University Hospital Magdeburg, Magdeburg, Germany.
Background:
Historically, [131I]I has been a common isotope in radionuclide therapy, with [177Lu]Lu-labelled radiopharmaceuticals now seeing a surge in use. These can include no-carrier-added or carrier-added [177Lu]Lu with slight impurities of [177mLu]Lu with a significantly longer half-life than [131I]I. Wastewater from therapy wards can contain a mixture of these radioisotopes. In some countries, national regulations require wastewater to be stored in dedicated systems before it is discharged into the public sewage system. To fulfill legal requirements, the nuclide specific activity concentration must be verified.
Purpose:
We evaluate a method for determining the activity concentration of [177mLu]Lu /[177Lu]Lu at equilibrium and [131I]I in pure and mixed samples in order to prove that the determined values are reliably below the limits for release.
Methods:
We analysed the emitted energy spectrum of 1 L samples with a wastewater counter using an energy window-based approach by evaluating measurements from two different time points. Based on the law of decay and the time and energy-dependent measured values, equation systems were set up to calculate the count rates for [131I]I and [177mLu]Lu, which were converted into activity concentration using calibration factors.
Results:
There is strong linear correlation between the nominal and determined activity concentrations (correlation coefficients R = 0.99; coefficient of determinations R2 = 0.99). We underestimate the actual activity concentration by a median of -1.4% for [177mLu]Lu and overestimate the activity concentration for [131I]I by a median of 7.1%.
Conclusion:
We show that an undercut of the clearance levels for material release is measurable. We analyse and determine activity concentrations of mixed samples consisting of [131I]I and [177mLu]Lu/[177Lu]Lu in equilibrium. The method is simple to implement using a conventional wastewater counter, however with a slightly increased effort, as two samples and measurements are required. The methodology can be adapted for the analysis of other nuclide mixtures.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:08Enhancing Efficiency and Radiolabeling Yields of Carbon-11 Radioligands for Clinical Research Using the Loop Method
Published on: December 20, 2024
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET