Analytical and Clinical Interference of Sample Hemolysis in Evaluating Blood Biochemical and Endocrine Parameters in
Dražen Kovačević1, Marko Cincović1, Mira Majkić1
1Faculty of Agriculture, University of Novi Sad, Square Dositeja Obradovića 7, 21000 Novi Sad, Serbia.
Abstract:
Hemolysis is a common cause of errors in laboratory tests as it affects blood parameters and leads to a positive or negative bias. This study aims to examine the relationship between the level of hemolysis (expressed as cell-free hemoglobin concentration, g/L) and the variability of metabolic and endocrine parameters and to determine the threshold level of hemolysis that causes an analytically and clinically significant bias for the twenty most frequently examined blood parameters in cows. Paired blood samples of 10 mL each were obtained from 30 cows. One was subjected to mechanical trauma and plasma was extracted directly from the other. Hemolyzed and non-hemolyzed samples from the same animal were mixed to obtain final samples with cell-free hemoglobin concentrations of 0, 1, 2, 4, 6, 8, and 10 g/L. Metabolic and endocrine parameters were measured in the samples and their deviation and the linear equation between the level of hemolysis and the deviation were determined. The following threshold values of hemolysis were determined, which correspond to the acceptable analytical (lower value) and clinical (upper value) levels of parameter variability: BHB 0.96 and 4.81; NEFA 0.39 and 3.31; GLU 0.38 and 3.90; ALB 1.12 and 6.11; TPROT 1.40 and 6.80; UREA 6.62 and 20.1; TBIL 0.75 and 5.65; AST 0.11 and 2.18; GGT 1.71 and 8.90, LDH 0.01 and 0.11, ALP 0.97 and 2.95; TGC 1.56 and 15.5; CHOL 1.29 and 8.56; Ca 5.68 and 25.7; P 0.57 and 8.43; Mg 1.10 and 8.47; INS 1.15 and 3.89; T3 8.19 and 15.6; T4 8.97 and 18.5; and CORT 2.78 and 11.22 g/L cell-free hemoglobin. Three decision levels are available for each metabolic and endocrine parameter: if hemolysis is below the lower (analytical) threshold value, results can be reported without restriction; if hemolysis is between the lower and upper thresholds, the results can be issued with guidance in the form of corrective linear equations; and if hemolysis is above the upper (clinical) threshold, the results and sample must be discarded. This method contributes to an optimal approach to hemolysis interference with metabolic profile parameters in blood samples from cows.
More Related Videos
11:12Ex Vivo Red Blood Cell Hemolysis Assay for the Evaluation of pH-responsive Endosomolytic Agents for Cytosolic Delivery of Biomacromolecular Drugs
Published on: March 9, 2013
08:29Measuring Liver Mitochondrial Oxygen Consumption and Proton Leak Kinetics to Estimate Mitochondrial Respiration in Holstein Dairy Cattle
Published on: November 30, 2018
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...