Accuracy of IVUS-Based Machine Learning Segmentation Assessment of Coronary Artery Dimensions and Balloon Sizing
Mitsuaki Matsumura1,2, Gary S Mintz1, Tomotaka Dohi2
1Clinical Trial Center, Cardiovascular Research Foundation, New York, New York, USA.
Background:
Accurate intravascular ultrasound (IVUS) measurements are important in IVUS-guided percutaneous coronary intervention optimization by choosing the appropriate device size and confirming stent expansion.
Objectives:
The purpose of this study was to assess the accuracy of machine learning (ML) automatic segmentation of coronary artery vessel and lumen dimensions and balloon sizing.
Methods:
Using expert analysis as the gold standard, ML segmentation of 60 MHz IVUS images was developed using 8,076 IVUS cross-sectional images from 234 patients, which were randomly split into training (83%) and validation (17%) data sets. The performance of ML segmentation was then evaluated using an independent test data set (437 images from 92 patients). The endpoints were the agreement rate between ML vs experts' measurements for appropriate balloon size selection, and lumen and acute stent areas. Appropriate balloon size was determined by rounding down from the mean vessel diameter or rounding up from the mean lumen diameter to the next balloon size. The difference of lumen area ≥0.5 mm2 was considered as clinically significant.
Results:
ML model segmentation correlated well with experts' segmentation for training data set with a correlation coefficient of 0.992 and 0.993 for lumen and vessel areas, respectively. The agreement rate in lumen and acute stent areas was 85.5% and 97.0%, respectively. The agreement rate for appropriate balloon size selection was 70.6% by vessel diameter only and 92.4% by adding lumen diameter.
Conclusions:
ML model IVUS segmentation measurements were well-correlated with those of experts and selected an appropriate balloon size in more than 90% of images.