The present work represents a reverse-phase high-performance liquid chromatography method in addition to stability studies for sequential estimation of remogliflozin etabonate, vildagliptin, and metformin HCl in tablet formulation.
Method:
The mentioned method utilizes a Phenomenex Luna C18 column (250×4.6mm, 5μm). It consists of a column oven's temperature of 35°C. Mobile phase includes a mixture of 50% phosphate buffer (pH - 6.8) and 50% acetonitrile along with a flow rate of 0.8mL/min and 20minutes of run time. The injection volume was 20μL; 217nm is a detection wavelength, and a PDA detector is used for detection.
Results:
The suggested technique was proven and validated per the ICH Q2 (R1) guideline. The combination was put under stress conditions that included acid, base, thermal, photolytic, and oxidative degradation. The combination was considerably degraded under oxidative, acidic, and basic circumstances for deterioration, and the degradation results were accurately identified from the observed peaks, demonstrating the method's effectiveness in detecting stability.
Conclusion:
The technique was quick, precise, sensitive, and accurate; as a result, it may be used in quality control laboratories and the pharmaceutical industry for routine quality monitoring of tablets containing all three medications.
More Related Videos
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Data Validation
Key parameters for method validation include:
Dipeptidyl Peptidase 4 Inhibitors