Verification of the effect of data-driven brain motion correction on PET imaging
Hayato Odagiri1,2, Hiroshi Watabe3, Kentaro Takanami2
1Department of Diagnostic Image Analysis, Tohoku University Graduate School of Medicine, Sendai, Miyagi, Japan.
Introduction:
Brain positron emission tomography/computed tomography (PET/CT) scans are useful for identifying the cause of dementia by evaluating glucose metabolism in the brain with F-18-fluorodeoxyglucose or Aβ deposition with F-18-florbetaben. However, since imaging time ranges from 10 to 30 minutes, movements during the examination might result in image artifacts, which interfere with diagnosis. To solve this problem, data-driven brain motion correction (DDBMC) techniques are capable of performing motion corrected reconstruction using highly accurate motion estimates with high temporal resolution. In this study, we investigated the effectiveness of DDBMC techniques on PET/CT images using a Hoffman phantom, involving continuous rotational and tilting motion, each expanded up to approximately 20 degrees.
Materials And Methods:
Listmode imaging was performed using a Hoffman phantom that reproduced rotational and tilting motions of the head. Brain motion correction processing was performed on the obtained data. Reconstructed images with and without brain motion correction processing were compared. Visual evaluations by a nuclear medicine specialist and quantitative parameters of images with correction and reference still images were compared.
Results:
Normalized Mean Squared Error (NMSE) results demonstrated the effectiveness of DDBMC in compensating for rotational and tilting motions during PET imaging. In Cases 1 and 2 involving rotational motion, NMSE decreased from 0.15-0.2 to approximately 0.01 with DDBMC, indicating a substantial reduction in differences from the reference image across various brain regions. In the Structural Similarity Index (SSIM), DDBMC improved it to above 0.96 Contrast assessment revealed notable improvements with DDBMC. In continuous rotational motion, % contrast increased from 42.4% to 73.5%, In tilting motion, % contrast increased from 52.3% to 64.5%, eliminating significant differences from the static reference image. These findings underscore the efficacy of DDBMC in enhancing image contrast and minimizing motion induced variations across different motion scenarios.
Conclusions:
DDBMC processing can effectively compensate for continuous rotational and tilting motion of the head during PET, with motion angles of approximately 20 degrees. However, a significant limitation of this study is the exclusive validation of the proposed method using a Hoffman phantom; its applicability to the human brain has not been investigated. Further research involving human subjects is necessary to assess the generalizability and reliability of the presented motion correction technique in real clinical scenarios.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
10:35Simultaneous PET/MRI Imaging During Mouse Cerebral Hypoxia-ischemia
Published on: September 20, 2015
Related Concept Videos
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...