Safety and performance of conduction system pacing: Real-world experience from a product surveillance registry
Pugazhendhi Vijayaraman1, Michael West2, Thomas Dresing3
1Geisinger Heart Institute, Wilkes-Barre, Pennsylvania.
Background:
Conduction system pacing (CSP), including His-bundle pacing (HBP) and left bundle branch area pacing (LBBAP), has been used as an alternative for pacemaker indicated patients requiring ventricular pacing.
Objective:
The purpose of this study was to characterize the safety and performance of HBP and LBBAP among patients enrolled in the Medtronic product surveillance registry (PSR).
Methods:
This observational analysis included patients who underwent pacemaker implantations for HBP or LBBAP with a Medtronic Model 3830 lead between January 2019 and December 2023 in the Medtronic PSR. The primary outcomes were lead-related complications and pacing capture threshold. Baseline characteristics, R-wave amplitude, impedance, and all-cause mortality were summarized.
Results:
A total of 2342 patients were included across 77 centers (mean age 74 years; 38.9% female). Of the patients implanted with a 3830 lead for CSP, 64.1% (n = 1502) had LBBAP placement and 35.9% (n = 840) had HBP placement. The most commonly reported indications for CSP were sinus node dysfunction (67.0%) and atrioventricular block (57.2%). LBBAP had lower pacing thresholds, higher R-wave sensing, and higher impedance (all P <.001) through 30 months. At 36 months postimplant, the lead complication rate for LBBAP and HBP was 2.5% and 6.3%, ,respectively with no difference in all-cause mortality.
Conclusion:
In a multicenter cohort of LBBAP and HBP patients treated with the catheter-delivered 3830 lead, lead-related complication rates were low and electrical parameters were stable through 30 months.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...