Technical note: Can photon-counting CT improve PET/CT's quantitative accuracy?
James Hamill1, Noah Birge1, Jim O'Doherty2,3
1Siemens Medical Solutions USA, Knoxville, Tennessee, USA.
Photon counting detector CT (PCD-CT) significantly improves the accuracy of linear attenuation coefficients (LACs) for positron emission tomography combined with computed tomography (PET/CT) imaging. This advancement enhances quantitative accuracy, particularly after iodine contrast administration.
Area of Science:
- Medical Imaging Physics
- Radiological Sciences
- Nuclear Medicine Technology
Background:
- Linear attenuation coefficients (LACs) in PET/CT are typically derived from energy-integrating detector CT (EID-CT), but are inaccurate with iodine contrast.
- Photon counting detector CT (PCD-CT) offers potential for improved LAC accuracy.
Purpose of the Study:
- To evaluate if PCD-CT can enhance the quantitative accuracy of PET/CT scans.
- To assess the impact of PCD-CT on LAC accuracy, especially in the presence of iodine contrast.
Main Methods:
- Two experiments were conducted: CT-only phantom studies and combined PET/CT phantom studies.
- EID-CT and PCD-CT were used to derive LACs from phantoms simulating human tissues, including bone and iodinated blood.
- PET images were reconstructed using LACs from both EID-CT and PCD-CT, with quantitative accuracy assessed via recovery coefficients (RC).
Main Results:
- PCD-CT-derived LACs showed a maximum error of 3%, compared to up to 18% error for EID-CT.
- PET quantitative accuracy improved with PCD-CT, indicated by more consistent recovery coefficients (mean RC 101.0% vs. 103.8% for EID-CT).
- Image consistency in both body and head orientations was enhanced using PCD-CT.
Conclusions:
- PCD-CT provides accurate LAC information for PET reconstruction within a single acquisition.
- This technology holds promise for improving the quantitative performance of PET/CT imaging.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET