Low-Power Long-Duration Versus High-Power Short-Duration Radiofrequency Ablation of the Atrioventricular Node
Sahil Zaveri1,2, Mahmoud Alsaiqali3, Howard Yu2
1From the Cardiovascular Research Program, VA New York Harbor Healthcare System, Brooklyn, NY.
Background:
Atrioventricular node (AVN) radiofrequency (RF) ablation is a highly effective treatment of atrial tachyarrhythmias that are resistant to other management modalities. To date, there is limited research that compares the properties of different RF ablation catheters. The current study aims to compare the effectiveness of several types of RF catheters in AVN ablation.
Methods:
A total of 66 patients, with a mean age of 73.27 years, underwent AVN RF ablation. The catheters used were categorized as unirrigated (UI), externally irrigated, and contact force sensing with 10 to 20 g of force. Externally-irrigated catheters were divided into 2 different settings: low-power long-duration (LPLD) (30 W, 45°C, and 60 seconds) and high-power short-duration (HPSD) (50 W, 43°C, and 12 seconds). We compared the success rate of the different RF catheters using logistic regression and lesion times using linear regression.
Results:
The distribution of the types of catheters used is UI in 48%, LPLD in 16%, and HPSD in 36% of patients. All ablation procedures were successful, with no immediate postprocedure complications. HPSD had a significantly shorter lesion time than UI catheters by 403.42 seconds (-631.67 to -175.17).
Conclusions:
UI catheters, LPLD, and HPSD were equally safe and effective in ablation procedures. The HPSD catheter had a significantly shorter lesion time and, thus, overall decreased procedure time.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias