Background:
MRI has been proposed as a new staging tool for bladder cancer (BC), but use is limited by its high costs and low availability. 29-MHz high-resolution micro-ultrasound (mUS) technology has been suggested as an alternative to detect BC and distinguish between muscle-invasive and non-muscle invasive BC.
Objective:
The aim was to compare the diagnostic accuracy of mUS vs. magnetic resonance imaging (MRI) in differentiating NMIBC and MIBC at definitive pathological examination.
Methods:
This is a prospective study of patients with a primary diagnosis of BC with either positive urine cytology (UC) or negative UC and a tumor size > 25 mm from a tertiary care high volume center. mUS, with the ExactVu system with an EV29L 29 MHz side-fire transducer, and a 3-Tesla MRI were performed before transurethral resection of bladder tumor (TURBT) in every patient before undergoing TURBT. We compared the imaging results with pathological reports.
Results:
The analyzed population consisted of 58 individuals. The reported mUS and MRI sensitivity, specificity, positive, and negative predictive values were 85.0%, 76.3%, 65.4%, and 90.6%, versus 85.0%, 50.0%, 47.2%, and 86.4%, respectively. In accuracy analysis, the AUC for mUS and MRI were respectively 0.807 and 0.675.
Conclusions:
In our population mUS seems to have a better performance in distinguishing NMIBC from MIBC. The main limitation of mUS is the probe shape that makes its use problematic in cases with a large prostate and inadequate rectal preparation. Further studies with a larger population are ongoing to compare and validate these techniques in this setting.