Related Experiment Video
Updated: Jun 21, 2025

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Angle Dependence of Electrode Lead-Related Artifacts in Single- and Dual-Energy Cardiac ECG-Gated CT Scanning-A
Piotr Tarkowski1, Elżbieta Siek2,3, Grzegorz Staśkiewicz2,3
1Department of Radiology, Medical University of Lublin, 20-090 Lublin, Poland.
Abstract:
Background: The electrodes of implantable cardiac devices (ICDs) may cause significant problems in cardiac computed tomography (CT) because they are a source of artifacts that obscure surrounding structures and possible pathology. There are a few million patients currently with ICDs, and some of these patients will require cardiac imaging due to coronary artery disease or problems with ICDs. Modern CT scanners can reduce some of the metal artifacts because of MAR software, but in some vendors, it does not work with ECG gating. Introduced in 2008, dual-energy CT scanners can generate virtual monoenergetic images (VMIs), which are much less susceptible to metal artifacts than standard CT images. Objective: This study aimed to evaluate if dual-energy CT can reduce metal artifacts caused by ICD leads by using VMIs. The second objective was to determine how the angle between the electrode and the plane of imaging affects the severity of the artifacts in three planes of imaging. Methods: A 3D-printed model was constructed to obtain a 0-90-degree field at 5-degree intervals between the electrode and each of the planes: axial, coronal, and sagittal. This electrode was scanned in dual-energy and single-energy protocols. VMIs with an energy of 40-140 keV with 10 keV intervals were reconstructed. The length of the two most extended artifacts originating from the tip of the electrode and 2 cm above it-at the point where the thick metallic defibrillating portion of the electrode begins-was measured. Results: For the sagittal plane, these observations were similar for both points of the ICDs that were used as the reference location. VMIs with an energy over 80 keV produce images with fewer artifacts than similar images obtained in the single-energy scanning mode. Conclusions: Virtual monoenergetic imaging techniques may reduce streak artifacts arising from ICD electrodes and improve the quality of the image. Increasing the angle of the electrode as well as the imaging plane can reduce artifacts. The angle between the electrode and the beam of X-rays can be increased by tilting the gantry of the scanner or lifting the upper body of the patient.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...

