Evaluation of a BioFire multiplex PCR panel for detection of joint infections using retrospective and prospectively

Angelica Moran1,2, Jekzaly Arellano1, Karen Bregman1

  • 1Department of Pathology and Laboratory Medicine, Endeavor Health, Evanston, Illinois, USA.

Insights

The new BioFire Joint Infection (JI) panel accurately detects bone and joint infection pathogens. While showing good agreement with standard cultures, potential false results necessitate clinical context for accurate diagnosis.

Area of Science:

  • Microbiology
  • Infectious Diseases
  • Molecular Diagnostics

Background:

  • Bone and joint infections require rapid and accurate pathogen identification for effective treatment.
  • Traditional culture methods can be time-consuming, delaying diagnosis and therapeutic intervention.
  • Multiplex PCR assays offer a faster alternative for detecting a broad range of pathogens.

Purpose of the Study:

  • To evaluate the performance of the FDA-approved BioFire Joint Infection (JI) panel for detecting pathogens in joint fluid specimens.
  • To compare the JI panel's results against standard bacterial culture methods.
  • To assess the diagnostic utility of the JI panel in a clinical setting.

Main Methods:

  • Retrospective testing of 63 frozen and prospective testing of 104 fresh residual joint fluid specimens using the BioFire JI panel.
  • Comparison of JI panel results with results from routine bacterial cultures (agar plates and blood cultures).
  • Calculation of positive percent agreement (PPA) and negative percent agreement (NPA) for both specimen types.

Main Results:

  • The JI panel demonstrated high PPA (92.8%) and NPA (97.1%) for frozen specimens.
  • For fresh specimens, PPA was 71.4% and NPA was 94.8%.
  • A total of 12 discrepancies (approximately 7%) were observed across 167 specimens tested.

Conclusions:

  • The BioFire JI panel shows good overall agreement with routine bacterial culture for diagnosing joint infections.
  • The panel has the potential to expedite diagnosis due to its rapid turnaround time.
  • Discrepancies highlight the importance of interpreting JI panel results within the appropriate clinical context to avoid potential false positives or negatives.