Application of metal artifact reduction algorithm in reducing metal artifacts in post-surgery pediatric low radiation
Jihang Sun1,2, Haoyan Li1, Tong Yu1
1Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China.
Background:
The commonly used methods for removing metal-induced beam hardening artifacts often rely on the use of high energy photons with either high tube voltage or high energy virtual monoenergetic images in dual-energy computed tomography (CT), the radiation dose was usually relatively high in order to generate adequate signals. This retrospective study is designed to evaluate the application of a metal artifact reduction (MAR) algorithm in reducing pedicle screw metal-caused beam hardening artifacts in post-surgery pediatric low radiation dose spine CT images.
Methods:
Seventy-seven children (3-15 years) who had undergone a low dose CT with 140 or 100 kV were enrolled. The radiation dose was 1.40 mGy for the 3-8 years old and 2.61 mGy for 9-15 years old children. There were 116 pedicle screws evaluated. The raw data were reconstructed with adaptive statistical iterative reconstruction-V (ASIR-V) at 50% strength, ASIR-V with MAR (AV-MAR), deep learning image reconstruction (DLIR) at high strength and DLIR with MAR (DL-MAR). The image quality concerning pedicle screws was evaluated objectively in terms of the length of beam-hardening artifact (LHA) and artifact index (AI), and subjectively using a 4-point scale (4 points: best, 3 points: acceptable).
Results:
Both AV-MAR and DL-MAR significantly reduced metal-induced beam hardening artifacts with smaller LHA (15.76±10.12 mm, a reduction of 57.24% and 15.66±10.49 mm, a reduction of 57.40%, respectively), and AI value (62.50±33.51, a reduction of 64.65% and 61.03±32.61, a reduction of 65.01%, respectively) compared to ASIR-V and DLIR (all P<0.01), The subjective image quality scores concerning the screws were 3.37±0.49 and 3.47±0.50 with AV-MAR and DL-MAR, respectively, higher than the respective value of 1.73±0.44 and 1.76±0.43 without MAR (all P<0.01).
Conclusions:
MAR significantly reduces the low-density artifacts caused by metal screws in post-surgery pediatric low-dose spine CT images, across different tube voltages, radiation dose levels and reconstruction algorithms. Combining DL-MAR further improves the overall image quality under low radiation dose conditions.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...