A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Comparative post-therapeutic dosimetry between 2D planar-based and hybrid-based methods for personalized Lu-177
Wuri Handayani1,2, Maythinee Chantadisai3, Benchamat Phromphao3
1Medical Physics Program, Department of Radiology, Faculty of Medicine, Chulalongkorn University, Bangkok, 10330, Thailand.
Planar dosimetry overestimates absorbed radiation dose in organs and tumors compared to hybrid methods. The hybrid approach offers better tumor visualization and organ delineation for 177Lu-therapies.
Area of Science:
- Nuclear Medicine
- Medical Physics
- Radiopharmaceutical Therapy
Background:
- Personalized dosimetry is crucial for optimizing radionuclide therapies like 177Lu-PSMA I&T and 177Lu-DOTATATE.
- Accurate absorbed dose calculation relies on appropriate imaging protocols and dosimetry software.
- HERMES HybridViewer offers advanced dosimetry capabilities integrating various imaging modalities.
Purpose of the Study:
- To compare absorbed doses in organs and tumors using different imaging protocols and calculation methods in HERMES HybridViewer.
- To evaluate dosimetry for 177Lu-PSMA I&T and 177Lu-DOTATATE therapies.
- To assess the impact of imaging methodology on radiation dose calculations.
Main Methods:
- Acquired whole-body planar and SPECT/CT images from 18 patients undergoing 177Lu-therapies.
- Calculated time-integrated activity (TIA) in organs and tumors.
- Determined absorbed doses using the Medical Internal Radiation Dose (MIRD) scheme via HERMES HybridViewer with OLINDA/EXM V.2.1 and NURBS digital phantoms.
Main Results:
- Planar-based dosimetry yielded higher dose per injected activity than hybrid-based dosimetry, attributed to organ overlap.
- The spleen showed the largest absorbed dose difference (up to 51.6%) between methods.
- Differences in other organs and tumors were less than 40%.
Conclusions:
- 2D planar dosimetry significantly overestimates absorbed dose compared to the hybrid method.
- The hybrid method provides superior tumor visualization and overlap-free organ delineation.
- Hybrid dosimetry is recommended for more accurate dose assessment in 177Lu-therapies.