Related Experiment Video
Updated: Jun 20, 2025

Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Changes of pulse wave transit time after haemodynamic manoeuvres in healthy adults: a prospective randomised
Johannes M Wirkus1, Fabienne Goss1, Matthias David2
1Department of Anaesthesiology, University Medical Center of the Johannes Gutenberg-University, Mainz, Germany.
Background:
Pulse wave transit time (PWTT) shows promise for monitoring intravascular fluid status intraoperatively. Presently, it is unknown how PWTT mirrors haemodynamic variables representing preload, inotropy, or afterload.
Methods:
PWTT was measured continuously in 24 adult volunteers. Stroke volume was assessed by transthoracic echocardiography. Volunteers underwent four randomly assigned manoeuvres: 'Stand-up' (decrease in preload), passive leg raise (increase in preload), a 'step-test' (adrenergic stimulation), and a 'Valsalva manoeuvre' (increase in intrathoracic pressure). Haemodynamic measurements were performed before and 1 and 5 min after completion of each manoeuvre. Correlations between PWTT and stroke volume were analysed using the Pearson correlation coefficient.
Results:
'Stand-up' caused an immediate increase in PWTT (mean change +55.9 ms, P-value <0.0001, 95% confidence interval 46.0-65.7) along with an increase in mean arterial pressure and heart rate and a drop in stroke volume (P-values <0.0001). Passive leg raise caused an immediate drop in PWTT (mean change -15.4 ms, P-value=0.0024, 95% confidence interval -25.2 to -5.5) along with a decrease in mean arterial pressure (P-value=0.0052) and an increase in stroke volume (P-value=0.001). After 1 min, a 'step-test' caused no significant change in PWTT measurements (P-value=0.5716) but an increase in mean arterial pressure and heart rate (P-values <0.0001), without changes in stroke volume (P-value=0.1770). After 5 min, however, PWTT had increased significantly (P-value <0.0001). Measurements after the Valsalva manoeuvre caused heterogeneous results.
Conclusion:
Noninvasive assessment of PWTT shows promise to register immediate preload changes in healthy adults. The clinical usefulness of PWTT may be hampered by late changes because of reasons different from fluid shifts.
Clinical Trial Registration:
German clinical trial register (DRKS, ID: DRKS00031978, https://www.drks.de/DRKS00031978).
Related Concept Videos
Pulse
The pulse serves as a clinical...
Special considerations while measuring pulse
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Regulation of Pulse
Blood Pressure
Pulse Assessment Sites