Minimal effect of scanning parameters on ultrasound shear wave elastography variability in tendons
Jennifer A Zellers1,2, Ran Li3, Rachana Vaidya1,2
1Program in Physical Therapy, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA.
Purpose:
Ultrasound shear wave elastography has potential use in assessing tendon tissue; however, reducing measurement variability remains challenging. The primary purpose of this study was to identify the amount of variability accounted for by ultrasound parameter (frequency, harmonics and CrossXBeam) settings on shear wave speed at two testing sites.
Methods:
Shear wave elastography images of the Achilles tendon were obtained from individuals with healthy tendons (n = 28) at two testing sites with standardised image acquisition/postprocessing protocols. Images were acquired at a range of frequencies (7-15 MHz) with CrossXBeam (a filtering technique) and harmonics settings toggled on and off. Variance decomposition analysis was performed to identify the amount of variability in shear wave speed accounted for by scan acquisition settings and testing sites.
Results:
Shear wave speed variance was primarily attributed to participants (56.87% of variance; residual error: 35%). All scanning parameters, testing site and interaction terms each accounted for less than 2.5% of the variance. A statistically significant, negative relationship was observed between shear wave speed and image quality (p = 0.001) suggesting poor image quality yields higher shear wave speed estimates.
Conclusion:
The findings of this study suggest that natural variation in Achilles tendon mechanics between individuals without tendon pathology accounts for most of the shear wave speed variability. Optimising image quality, which may be observed in higher frequencies, should be considered to improve shear wave speed estimation. Clinically, this study highlights the need to take multiple images, maintain consistent ultrasound settings when tracking patient progress over time and use caution when comparing raw values from tendon scans performed in different clinics with shear wave elastography.
Level Of Evidence:
Level III.