Recent Recall of Iron Reagent-Investigation of Potential Reagent Contamination and Assay Improvement Strategy
Madhusudhanan Narasimhan1, Kefyalew Jaleta2, Shishir Adhikari2
1Department of Pathology, University of Texas Southwestern Medical Center, Dallas, TX, United States.
Background:
Recently, a major manufacturer recalled several lots of iron assay reagent due to positive bias of roughly 15%-30% and the cause remains unknown. This study investigated the root cause of this positive bias and evaluated a simple practical approach to improve the assay.
Methods:
Performance comparison of recalled and unimpacted iron assay kits was done utilizing calibrators, quality control (QC) materials, and 42 remnant patient samples. Spectral scan and trace elements analysis of R1 and R2 reagents was performed. Copper (Cu) and thiourea (TU) spiking experiments were utilized to elucidate the cause and prevention of positive bias seen with recalled lots.
Results:
Iron measurements in QC materials and patient samples using recalled reagents generated a positive bias of 17.5% and 21%, respectively. Correspondingly, the recalled R2 reagents, but not R1, showed a rise in basal absorbance along with an unanticipated presence of Cu (22.7 µg/dL) and lead (7.5 µg/L). Cu spiking to recalled and unimpacted R2 reagent intensified the reagent color besides falsely increasing its absorbance, calibration factor, and patient iron measurements. Interestingly, addition of TU (65 mmol/L) to R2 reagent from unimpacted lot prevented the short-term and prolonged Cu-induced spurious rise in calibration factor and patient iron estimations.
Conclusions:
We conclude that accidental copper contamination of R2 reagent during manufacturing could be a reason underlying the positive bias in the recalled iron reagent lots. Addition of TU in ferene-containing R2 reagent is a simple and effective means to prevent Cu-induced false elevation in iron values.
More Related Videos
05:36Author Spotlight: Assessing the Impact of Novel Iron Chelators on Cancer Cell Metabolism
Published on: February 23, 2024
04:48Setup of Capillary Electrophoresis-Inductively Coupled Plasma Mass Spectrometry CE-ICP-MS for Quantification of Iron Redox Species FeII, FeIII
Published on: May 4, 2020
Related Concept Videos
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Precipitation and Co-precipitation
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...