Time-of-flight PET/CT suppresses CT based attenuation correction and scatter coincidence correction errors due to
Yuya Watanabe1, Shota Hosokawa1, Yasuyuki Takahashi1
1Hirosaki University Graduate School of Health Sciences, Hirosaki-shi, Aomori, Japan.
Objectives:
This study aimed to examine the influence of changes in CT values on PET images, specifically focusing on errors in CT-based attenuation correction and scatter coincidence correction (CTAC/SC) caused by gastrointestinal gas. Furthermore, it aimed to demonstrate the effectiveness of time-of-flight (TOF) PET in reducing CTAC/SC errors.
Methods:
PET images were reconstructed using multiple CT images with varying CT values. The study then compared the fluctuations in pixel values of the PET images corresponding to the different CT values utilized for CTAC/SC between non-TOF and TOF acquisitions.
Results:
PET pixel values fluctuated with changes in CT values. In the phantom study, TOF showed a significantly smaller change in PET pixel value of 1.00±0.27 kBq/mL compared to 3.72±1.33 kBq/mL in the non-TOF at sites with a CT change of +1000 HU. In the patient study, a linear regression analysis was performed to determine the effect of changes in CT values due to gastrointestinal gas migration on standard uptake value (SUV).The results showed that the TOF group had a lower ratio of change in SUV to change in CT values compared to the non-TOF group. These findings revealed that PET pixel values exhibited fluctuations in response to changes in CT values, and TOF-PET effectively mitigated CTAC/SC errors arising from gastrointestinal gas.
Conclusions:
TOF-PET has the potential to reduce the occurrence of suspicious accumulation.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...