Appropriate and inappropriate ICD shocks in patients with LVADs: Prevalence, associated factors, and etiologies
Andrew Andreae1, Eric Black-Maier2, Kelly Arps1
1Division of Electrophysiology, Duke University Medical Center, Durham, North Carolina.
Background:
Implantable cardioverter-defibrillator (ICD) shocks are a common complication after left ventricular assist device (LVAD) implantation; however, data on their frequency and causes are limited.
Objective:
The purpose of this study was to define the incidence, programming, patient characteristics, and factors associated with appropriate and inappropriate ICD shocks in persons with LVADs.
Methods:
We performed a retrospective review at Duke University Hospital of all LVAD recipients implanted between January 1, 2013, to June 30, 2019, with a preexisting ICD. ICD shocks were adjudicated by the treating physician and a second reviewer for the purpose of this study.
Results:
Among 421 patients with an ICD in situ undergoing LVAD implant, 147 (33.9%) had at least 1 shock after LVAD implantation. Among 134 patients with complete device history, a total of 330 shock episodes occurred: 255 (77.3%) appropriate and 75 (22.7%) inappropriate. Etiologies for inappropriate shocks included supraventricular tachycardia (n = 66 [20.0%]), physiological oversensing (n = 1 [0.3%]), and nonphysiological oversensing (n = 8 [2.4%]) including LVAD electromagnetic interference (n = 1 [0.3%]). ICD programming with shorter detection delay (P <.001) and absence of antitachycardia pacing programming (P = .001) in high-rate zones was seen more commonly in inappropriate shock than appropriate shock.
Conclusions:
The rate of inappropriate shocks in LVAD recipients is very high and most often is due to supraventricular arrhythmias. LVAD electromagnetic interference is a rare cause of ICD shock. Implementation of current consensus American Heart Association recommendations for LVAD programming with long detection delays and high rate cutoffs may help prevent inappropriate ICD shocks.
More Related Videos
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Blood Pressure Imbalances and Circulatory Shock
Blood Pressure: Hypertension and Hypotension
Normal blood pressure is 120/80 mm Hg. Elevated blood pressure is 120-129/under 80 mm Hg. Hypertension, warranting treatment at 130/80 mm Hg, is often asymptomatic and can lead to severe cardiovascular events, aneurysms, peripheral arterial disease, chronic renal disease, or cardiac...
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
Antiarrhythmic Drugs: Class I Agents as Sodium Channel Blockers
Class 1A Antiarrhythmic Drugs: These drugs work by moderately blocking sodium channels,...