Related Experiment Video
Updated: Jun 18, 2025

Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Diagnostic Performance of Point-of-Care High-Sensitivity Troponin in the Exclusion of Non-ST-Elevation Myocardial
Daniel Zalama-Sánchez1, Carlos Del Pozo Vegas2,3, Ancor Sanz-García4,5,6
1Emergency Department, Hospital Universitario Virgen de la Victoria, 29010 Málaga, Spain.
Background:
This study evaluates the diagnostic performance of high-sensitivity troponin using point-of-care testing (POCT) devices compared with main laboratory measurements for ruling out non-ST-elevation myocardial infarction (NSTEMI) in emergency department (ED) patients presenting with non-traumatic chest pain.
Methods:
This multicenter, observational, prospective, non-interventional study was conducted in two Spanish hospitals from 1 June to 31 December 2023 and included adult patients presenting with non-traumatic chest pain admitted to the ED. High-sensitivity troponin levels were measured using both the Siemens Atellica® VTLi POCT device and main laboratory testing, with data collected on analytical results and measurement times.
Results:
Of the 201 patients who met the inclusion criteria, a significant correlation was observed between the POCT and laboratory assays. The area under the curve (AUC) of the ROC curve was consistently greater than 0.9, indicating a high diagnostic accuracy for ruling out NSTEMI. In addition, measurement times were significantly reduced using POCT compared to the core laboratory.
Conclusion:
These results suggest that high-sensitivity troponin POCT devices offer comparable diagnostic performance to traditional laboratory methods for the diagnosis of NSTEMI in the emergency department, potentially speeding up clinical decisions and optimizing resource utilization.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...