An HPLC-UV Method to Assess Human Plasma 25(OH)D3
Alexandra Tijerina1, Aurora Garza2, Abad López1
1Faculty of Public Health and Nutrition, Autonomous University of Nuevo Leon, Monterrey 64460, NL, Mexico.
This study validated an HPLC-UV method for assessing vitamin D status by accurately quantifying 25-hydroxyvitamin D3. The reliable method demonstrated excellent linearity and recovery rates for clinical use.
Area of Science:
- Analytical Chemistry
- Biochemistry
- Clinical Diagnostics
Background:
- Accurate assessment of vitamin D status is crucial for public health.
- 25-hydroxyvitamin D3 (25(OH)D3) is the primary indicator of vitamin D status.
- Validated analytical methods are essential for reliable vitamin D testing.
Purpose of the Study:
- To validate a High-Performance Liquid Chromatography-Ultraviolet (HPLC-UV) method for quantifying 25-hydroxyvitamin D3 (25(OH)D3).
- To establish the method's linearity, precision, accuracy, and robustness for assessing vitamin D status.
Main Methods:
- HPLC-UV analysis using an Acclaim™ 120 C18 column.
- Preparation of 25(OH)D3 calibration curves from 5 to 50 ng/mL.
- Extraction of vitamin D from plasma samples using acetonitrile and formic acid.
- Validation conducted according to FDA guidelines.
Main Results:
- Excellent linearity for the 25(OH)D3 calibration curve (R² = 0.9989).
- Detection and quantification limits of 1.1703 ng/mL and 3.5462 ng/mL, respectively.
- High percentage recovery rates (92.2% to 97.1%) indicating method accuracy.
Conclusions:
- The validated HPLC-UV method is reliable for quantifying 25(OH)D3 in human plasma.
- This method can be effectively used to assess vitamin D status.
- The method's robustness was confirmed through analysis of temperature and flow rate variations.