Related Experiment Video
Updated: Jun 18, 2025

Transaxillary First Rib Resection for Treatment of the Thoracic Outlet Syndrome
Published on: September 13, 2020
Comparison of Wireless Continuous Axillary and Core Temperature Measurement after Major Surgery
Anders Blom Nathansen1, Jesper Mølgaard1, Christian Sylvest Meyhoff2,3
1Department of Anesthesiology, Center for Cancer and Organ Diseases, Rigshospitalet, University of Copenhagen, Blegdamsvej 9, 2200 Copenhagen, Denmark.
Background:
Temperature is considered one of the primary vital signs for detection of complications such as infections. Continuous wireless real-time axillary temperature monitoring is technologically feasible at the general ward, but no clinical validation studies exist.
Methods:
This study compared axillary temperature with a urinary bladder thermometer in 40 major abdominal postoperative patients. The primary outcome was changes in axillary temperature registrations. Secondary outcomes were mean bias between the urinary bladder and the axillary temperatures. Intermittent frontal and tympanic temperature recordings were also collected.
Results:
Forty patients were monitored for 50 min with an average core temperature of 36.8 °C. The mean bias was -1.0 °C (LoA -1.9 to -0) after 5 min, and -0.8 °C (LoA -1.6 to -0.1) after 10 min when comparing the axillary temperature with the urinary bladder temperature. After 20 min, the mean bias was -0.6 °C (LoA -1.3-0.1). During upper arm abduction, the axilla temperature was reduced to -1.6 °C (LoA -2.9 to -0.3) within 1 min. Temporal skin temperature measurement had a resulted in a mean bias of -0.1 °C (LOA -1.1 to -1.0) compared with central temperature. Compared with the mean tympanic temperature, it was -0.1 °C (LoA -0.9 to -1.0) lower than the urinay bladder temperature.
Conclusions:
Axillary temperature increased with time, reaching a mean bias of 1 °C between axillary and core temperature within 5 min. Opening the axillary resulted in rapidly lower temperature recordings. These findings may aid in use and designing corrections for continuous axillary temperature monitoring.
More Related Videos
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
08:50Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management
Published on: September 2, 2015
Related Concept Videos
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...